Body Infrared Thermometer
User’s Manual

Please read this manual before use!
Thank you for choosing our Infrared Thermometer. The PC868 Infrared Thermometer is used to measure the temperature of an object based on the relationship between temperature and measurable infrared radiation. Simply aim the unit’s probe toward the surface to be measured to obtain a quick and accurate temperature. To ensure proper use, please be sure to read this user manual carefully, paying close attention to the safety precautions.
All objects above absolute zero temperature emit a certain percentage of infrared radiation energy based on their temperature. The amount of the radiation energy and the distribution of the wavelength have very closely related. When the human forehead’s temperature is 36-37℃, it emits wavelength 9-13um of infrared radiation. Based on this principle, according to the relationship between surface forehead temperature and human forehead temperature, we are able to measure the human forehead’s actual temperature by measuring surface forehead temperature.
Warning 
Notice 

Recommendations 
This thermometer is intended with non-contact to measure forehead temperature at home or hospital, including anyone, such as infants, children, and adults.
For safety reasons, children’s or the baby’s temperature must be measured by parents or adults.
The infrared thermometer has the following measurement mode:
Normal temperature range for different measuring position
| Measuring position | Normal temperature (℃) | Normal temperature (℉) |
| Anus | 36.6-38.0 | 97.9-100.4 |
| Oral | 35.5-37.5 | 95.9-99.5 |
| Armpit | 34.7-37.3 | 94.5-99.1 |
| Forehead | 35.8-38.0 | 96.4-100.4 |
Normal forehead temperature range for different ages
| Ages | Normal temperature (℃) | Normal temperature (℉) |
| 0-2 years old | 36.4-38.0 | 97.5-100.4 |
| 3-10 years old | 36.1-37.8 | 97.0-100.0 |
| 11-65 years old | 35.9-37.6 | 96.6-99.7 |
| >65 years old | 35.8-37.5 | 96.4-99.5 |
NOTE: The normal temperature and difference between the different body parts are individual. The define yours, measure your temperature for a least 2 weeks at the same ear canal, forehead position, and time.
NOTE: When consulting your physician, communicate that the infrared thermometer PC868 temperature is a temperature measured which a position, note the individual’s normal infrared thermometer PC868 temperature range as an additional reference.
NOTE: Because the forehead temperature is affected obviously by the external environment (eg: environment, air convection and skin tone, etc), we advise that you take the forehead temperature only as a reference.

| 1. IR sensor | 2. LCD display |
| 3. Mode selection | 4. Down button |
| 5. UP button | 6. Mode button |
| 7. Measurement Trigger | 8. Battery cover |
【LCD display description】

| 1. Surface mode symbol 2. Body mode symbol 3. Digital readout 4. Battery symbol |
5. Memory symbol 6. Save data readout 7. Temp. °C(Celsius)/ °F (Fahrenheit) scale 8. Buzzer symbol |
【Preparation】
【Instruction for use】
Advice for the first use
For stable and reliable results, it is essential to check the infrared thermometer and changes as needed, as follows:
| 1st step: | Take the temperature of a person using a conventional thermometer, you will get 37.5℃ (99.5℉) for instance. |
| 2nd step: | Take the temperature of the same person using the device keeping the 3 to 5cm distance between the thermometer and forehead (take care to remove any obstacle which could alter the measurement (hair, perspiration ….). If you get 37.5 ℃(99.5℉), the device is properly set and ready for use. If you get a lower temperature such as 36.4 ℃(97.5℉), your difference is 1.1℃(2.2℉). You should adjust the temperature on the device and add the difference, i.e.1.1℃(2.2℉). To do it, press the MODE button for 2 seconds, the screen displays F1, press MODE button again until you get F3, press the UP button in order to add the difference (in our example 1.1 ℃ (2.2℉). |
| 3rd step: | To check, take the temperature again using the product. |


| Mode code | Function | Operation step | |||
| 1 | F1 | Set ℃/℉ scale | Press the MODE button for 2 seconds, the screen will display F1 on the bottom right, press UP or DOWN button can change the temperature unit for℃or℉. | ||
| 2 | F2 | Set alarm temp. | Press the MODE button for 2 seconds, the screen display F1, press MODE button to get F2, select UP to increase the threshold by 0.1℃ (0.1℉), DOWN to reduce it by 0.1℃(0.1℉). The alarm threshold default value is 38.0℃(100.4℉). | ||
| 3 | F3 | Set difference | Press the MODE button for 2 seconds, the screen display F1, press two times to get F3, select UP to increase the difference by 0.1℃(0.1℉), DOWN to reduce it by 0.1℃(0.1℉). In the case of seasonal or environmental changes, verification and adjustment should be carried out. Note: This function is only an effective body. |
||
| 4 | F4 | Set buzzer on/off | Press the MODE button for 2 seconds, the screen display F1, press 3 times to get F4. Select UP to turn on the buzzer, press DOWN to stop it. | ||
| 5 | Measurement mode | There are 2 measurement modes for this device. That is the forehead, surface. It is able to take the body temperature of human beings, take the temperature of an area or an object, food, a liquid, or a room. | |||
| Forehead mode | Put the mode selection button on the left, measurement range 32.0 to 43.0℃(86.0 to 109.4℉). | ||||
| Surface mode | Put the mode selection button on the right, measurement range 0 to 100℃(32.0 to 212.0℉). | ||||
| 6 | 3 colors backlight | In body mode, the device is able to judge whether a fever or not by a different color of backlight: 32.0 to 37.4℃(86.0 to 99.3℉) – Green 37.5 to 37.9℃(99.5 to 100.2℉) – Orange 38.0 to 43.0℃(100.4 to 109.4℉) – Red In surface mode, the backlight always keeps green. |
|||
| 7 | Data memory | Date memory automatically after temperature measurements, which will display at the right corner of the LCD. Press the UP or DOWN button to display the last temperature measurement. | |||
— Range of alarm temperature
When into the “F-2” mode, the range of settable alarm temperature is from 37.6℃(99.7℉) to 43.0℃(109.4℉).
— Range of temperature difference
When into the “F-3” mode, the range of settable temperature difference is from 5℃(9℉) to -5℃(-9℉).
Note: F-3 mode is only for professional doctor and nurse use!
Temperature measuring
–Taking Temperature in Forehead Temperature Mode (Adjusted Mode)
– Before taking the temperature, please fix the forehead hair to prevent measured deviation.
– The sweat on the head or cosmetic can affect the accuracy of measurement, please maintain the cleanness of the forehead when measuring.
– It is normal that there may be temperature difference depending on various skin types and color since different skin type will reflect different voltage of infrared ray.
– Don’t use the thermometer outdoors.
NOTE:
–Taking Temperature in Surface Temperature Mode (Direct Mode)
Automatic shutdown
If you do nothing after the temperature measurement is completed and 15 seconds later, the thermometer will automatically shut down.
Memory function
In the power-on state, press the Up/Down button into memory query mode, the LCD will be showing the recorded values or “ —-“ symbol if not recorded accompanied with memory M symbol. The PC868 can restore up to 99 sets of data. the query interface can display memory serial number, body mode, and recorded temperature values.
Remark: The temperature result of surface mode can not be saved. And the abnormal body temperature result can not be saved too (such as Lo, Err, or Hi).
Backlight unction
The backlight will be according to the different measuring temperature to lighting different color. The detailed as below table:
| Measuring temperature | Backlight | Lighting time |
| < 37.5℃ | Green backlight | 7s |
| 37.5 to 37.9℃ | Orange backlight | 7s |
| ≥38.0℃ | Red backlight | 7s |
| Remarks: This specification of this table is only for the factory default state. If the user into the setup mode to set the orange or red backing temperature, then will compliance with the setting result. | ||
Restore the factory default function
Press the [Setup] button for 5s to restore the factory default status. The factory default setup is “℃” and “Body measurement” mode.
Battery installment and replacement
Notice:
To protect the environment, dispose of empty batteries at your retail store or at appropriate collection sites according to national or local regulations.
NOTES:
The thermometer is initially calibrated at the time of manufacture.
If this thermometer is used according to the user’s instruction, periodic re-adjustment is not required. If any time your question the accuracy of measurement, please contact distributor or manufacturer, the contact information sees the last page.
Only use original accessories. Check that the contents of the delivery are complete.
| Quantity | Parts |
| 1pc | PC868 device |
| 2pcs | AAA battery |
| 1pc | User Manual |
| Troubles or error message | Checklists or situation | Countermeasures or solution |
| No response/Automatically reset | The batteries are used up? | Replace new batteries. |
| The battery in the wrong polarity or type? | Take out the batteries and replace new ones. Take out batteries and reinsert them correctly. |
|
| Poor battery contact | ||
| The thermometer shows the symbol “Hi” | Temperature hampered by an air flux. | Please leave the status and wait for 30 minutes to measure. Re-measure according to the manual. |
| In the forehead measurement mode: –Temperature readings are too close together. — Measured the other object, such as the sunlight, the air from the fireplace. Hi: Higher than 43.0℃; |
||
| In the object measurement mode: –Temperature readings are too close together. — The object temperature is higher than 100.0℃. Hi: Higher than 100.0℃; |
||
| The thermometer shows the symbol “Lo” | The hair and sweat prevent the temperature achievement. | |
| Temperature hampered by an air flux. | ||
| In the forehead measurement mode: — The measuring distance is too far. — Measured the other object, such as the air from the air conditioner. Lo: Less than 32.0℃ |
||
| In the object measurement mode: — The measuring distance is too far. — Have water vapor condenses on the lens. Lo: Less than 0℃ |
||
![]() |
The ambient temperature is beyond of range of measurement (10℃-40℃) | Keep the thermometer in the room Whose Temperature is (10℃-40℃) for 30 minutes |
| The sensor or hardware is damaged | Excluding the possibility of temperature allowance first, then send the device to your dealer for repair | |
![]() |
Lower battery, however, you can’t use it | Replace the new battery. |
| Device name | Infrared Thermometer |
| Model | PC868 |
| Measurement mode | Forehead and object temperature modes |
| Power supply | d.c.3V, 2 pcs AAA battery |
| Measuring range: | For forehead temperature: 32.0-43.0℃ For object surface temperature: 0-100℃ |
| Measuring accuracy: (At laboratory conditions) | for forehead temperature: ± 0.2 during 35. 5℃ -42.0℃; ± 0.3 during 32.0℃ -35.4℃ and 42.1℃ -43.0℃ |
| Clinical repeatability: | Within ±0.3℃ |
| Resolution of display | 0.1℃/0.1℉ |
| Operation condition | 10-40℃, Relative humidity 15-85%, 70-106KPa |
| Storage condition | -25-50℃, Relative humidity ≤95%, 70-106KPa |
| Size | 149(L)×95(W)×45(H)mm |
| Weight | Approx. 145g(including battery) |
| High body temperature hint | ≥38℃ |
| Grade of waterproof | IP22 |
| Electric shock | Internally powered ME equipment |
| Applied part | Type BF applied part, including the whole unit |
| Mode of operation | Continuous operation |
| Product life | 5 years |
| Software version | V1.0 |
| Note: Not intended to be sterilized. Not for use in an OXYGEN RICH ENVIRONMENT |
|
*The above specifications are subject to change without prior notice.
Shenzhen PACOM Medical Instruments Co., Ltd declares that the PC868 complies with the following applicable standards:
| EN 980 | Symbols for use in the labeling of medical devices |
| EN 1041 | Information supplied by the manufacturer with medical devices |
| EN 60601-1 | Medical electrical equipment Part 1: General requirements for basic safety and essential performance Medical electrical equipment — Part 1-2: General |
| EN 60601-1-2 | requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests Medical electrical equipment ‒ Part1-6: General |
| EN 60601-1-6 | requirements for basic safety and essential performance ‒ Collateral standard: Usability |
| EN 60601-1-11 | Medical electrical equipment ‒ Part 1-11: General requirements for basic safety and essential performance ‒ Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in home healthcare environment Clinical thermometers ‒ Part 5: Performance of infra-red |
| EN 12470-5 | ear thermometers (with maximum device) Medical electrical equipment part 2-56: particular |
| ISO 80601-2-56 | requirements for basic safety and essential performance of clinical thermometers for body temperature measurement |
| EN 62304 | Medical device software – Software life-cycle processes Medical devices ‒ Application of usability engineering to |
| EN 62366 | medical devicesBiological evaluation of medical devices – Part 1: Evaluation |
| EN ISO 10993-1 | and testing within a risk management process |
Dispose of the device in accordance with the regulation applicable at the place of operation. Dispose of at public collection point in the EU countries ‒ 2002/96/EC WEEE Directive.
If you have any queries, please refer to the local authorities responsible for waste disposal.
NOTES:
To protect the environment, dispose of empty batteries at your retail store or at appropriate collection sites according to national or local regulations.
Dispose of at public collection point in the EU countries ‒ 2006/66/EC Directive.
![]() |
Follow operating instructions |
![]() |
BF type applied part |
![]() |
Disposal in accordance with Directive 2002/97/EC (WEEE) |
![]() |
Complies with the European Medical Device Directive (93/42/EEC and amended Directive 2007/47/EC. Notified Body is SGS FIMKO YO. |
| Manufacturer information: The manufacturer is: Shenzhen PACOM Medical Instruments Co., Ltd. |
|
![]() |
Authorized Representative in the European Community. The European representative is: Shanghai International Holding Corp. GmbH (Europe) |
IP22 |
IP code of the device: this device’s grade of against ingress of solid foreign objects — ≧ 12.5mm diameter (and the against access to hazardous parts with finger); the grade of waterproof is dripping (150 tilted). |
![]() |
Batch code |
| Date of manufacture |
We grant you a lifetime guarantee after the date of purchase for one year. Any damage caused by improper handling shall not be covered by the guarantee. Batter and packaging are also excluded from the guarantee. All other damage claims are excluded. A guarantee claim must be submitted with the purchase receipt. Please pack your defective instrument well and send it with sufficient postage to the distributor.
Shenzhen Pacom Medical Instruments Co., Ltd.
Address: On the 8th floor of B District, B Building, No.5, Industry
five road, Jiangbian Community, Songgang,
Shenzhen, China.
Tel:86-755-32920339
Website: www.pacomsz.com
Shanghai International Holding Corp. GmbH (Europe)
Address: Eiffestrasse 80, 20537 Hamburg Germany
Tel: 0049-40-2513175
Fax: 0049-40-255726
E-mail: [email protected]
INFRARED
THERMOMETER
Instruction Manual
ModeLASAP-D02
Manufacturer: ASAP Technology(Jiangxi)Co.,Ltd
Add: West District Industrial Park,343100 JI’ AN TOWN, Jiangxi, China Factory
Add: West District Industrial Park,343100 .11′ AN TOWN, Jiangxi, China
Tel: +86 796 7118988
version: vl. 1
This product is a professional infrared thermometer for measuring temperature. It is widely used in all industries. It is easy to use, with Mode selection, LCD display, Buzzer reminder, Memory access, Backlight selection, Temperature transformation setting, Alarm threshold setting, Automatic shutdown functions, etc.
[Product parameter Major structure]
Product parameter
Temperature measuring range: 32.0 °C-42.9°C
Temperature measuring time: about 1 second
The best measuring distance: 1-5cm
Display mode: LCD screen display
Product size: 154x96X42mm
Product weight: C93g
This infrared thermometer has an automatic shutdown and self-detection function. Current consumption: OFF 10µA, ON‘.100mA
Power: 2xAA battery(3.0V)
Major structure
It is made of ABS plastic housing, cup copper head construction, copper head construction, lens, PCB circuit board, IC, electrical resistance, electric capacity, infrared sensor, LCD, buzzer and battery. [Usage range) It is used to measure the temperature of all kinds of objects.
Getting a better knowledge of the principle of infrared temperature measurement can help you to use this product correctly and make the test data accurately.

As shown in (Fig3-1),the appearance of the whole machine is composed of the following parts:
|
|

| Measuring way | Non-contact infrared temperature measurement | |
| Measuring range | 32°C-42.9°C | |
| Display accuracy | 0.1°C | |
| Test accuracy | 32.0°C-34.9°C | ±0.3°C |
| 35.0°C-42.0°C | ± 0.2°C | |
| 42.1°C-42.9°C | ±0.3°C | |
| Usage environment | Environment temperature 16″0–35^C /Relative humidity C.80% / Atmospheric pressure 860h Pa-1060h Pa |
|
| Transport and storage environment |
Environment temperature-20°C—+50°C / Relative humidity-.C.85% / Atmospheric pressure 80KPa— 106KPa |
| Power | DV3V(2AA batteries) |
| Display unit | Centigrade degree(°C)/Fahrenheit degree(°F) |
| Automatic shutdown | About 8 seconds |
| Power reminder | 2.5V±0.2V |
| Backlight | Three color backlight |
| Memory group number | 32 groups for human and object temperature data each |
| Weight | C 93g |

1) Buttons
Measuring button: short press to measure/power on
Up button(+): up data button to check forward record, alarm value setting, fine-tune Settings, buzzer, measuring unit setting, delete current memory
Down button(-): down data button to check the backward record, alarm value setting, fine-tune Settings, buzzer, measuring unit setting
Mode button: object/human
Setting button: setting enter
2) Measuring human’ s temperature
3) Measuring object’s temperature

• Overtemperature Mode:
| Menu | Function | Down button | Up button | Initial value | Remark |
| F-1 | Temperature unit setting |
Temperature unit switching |
Temperature unit switching |
Centigrade degree |
Can choose with memory function |
| F-2 | Alarm setting | Reduce 0.1 degree |
Increase 0.1 degree |
38 degrees | The objects’ mode
is invalid; valid range |
| F-3 | Temperature transformation |
Reduce 0.1 degree |
Increase 0.1 degree |
0.0 degree | The objects’ mode isinvalid;valid range +_ 1.6 degrees |
| F-4 | Buzzer setting | On Off switch’setting |
On, Off switch setting |
On | Can choose with memory function |
Long press the Setting button for 2 seconds and short press the Setting button once Fl appears on the display screen. It enters F2 alarm setting after the display screen shows F2. The current initial value is 38.0 °. Press the down button to shift down by 0.1 °, and the up button to shift up by 0.1 ° . Press the Setting button to confirm and directly enter F3 temperature transformation setting.
Long press the Setting button for 2 seconds or short press the Setting button for two times to enter the F3 setting. The current initial value of the temperature migration interface is 0.0 degree. Press the down button to reduce 0.1 degree and the up button to increase 0.1 degree. Press the Setting button to confirm and directly enter F4 buzzer setting.
Long press the Setting button for 2 seconds or short press Setting button for 3 times to enter the F4 buzzer setting. Current initial value of buzzer is on, press down/up button to switch the On and Off. Choose “on” or “off” on the display screen and confirm by pressing the Setting button. If do not want to save the Settings, do not need to press the mode setting button to save, wait 8 seconds before entering the sleep mode to exit.
After each measurement, the instrument will record the test data automatically and display it in the lower right corner of the screen. Press “up” and “down” button to view the stored value. There’ re 32 group numbers about both for human body’ s temperature and object’ s temperature. It will automatically overwritten when exceed 32 group numbers.
After short pressing the measuring button, press it to start to wake up. At the same time, the screen will be fully displayed (500ms) and the temperature will be measured.
After Power on, long press “+” button for 3 seconds, when showing “CL?’ on screen means memory record has been deleted.
| Error message | Reason | Handing method |
| HI | Human body, s ternperature >42.9°C Or object’ s temperatu re >80°C Or the environment’s temperature exceeds the maximum temperature range |
Please use it within the measurable temperature range. If can’ t fix, please contact after-sales service. |
| LO | Human body’ s temperature <32.0°C Or object’ s temperature <0 °C Or |
Please use it within the measurable temperature range. If can’t fix, please contact after-sales service. |
| the environment s temperature exceeds the minimum temperature range | ||
| Err | Data error | Please contact after-sales service |
| Battery icon flash | Battery low warning | Please change battery |
Infrared thermometer is a precision electronic product, please use with care and attention:
Product Warranty Card
From the date after purchasing, free warranty for one year (two years maintenance fee will take into consideration the material artificial fee)
Our company don’t provide free warranty service for the following failures which caused by users’ personal reasonsAs follows:
When asked free services, please present the valid warranty card and shopping vouchers. When asked warranty service, please bring the product to the point of sale for repair. Charge will be effected according to the regulations outside the scope of the warranty repair service.
Limited Warranty
Warranty Card
Purchase Date:…..
Distributor:………
Series Number:…… Store stamp
Certificate of Approvel
Model: ASAP-D02
Production Date: ….
Inspector:….
The product has been tested and proved to be qualified for delivery.
]]>

DEVICE DESCRIPTION AND INTENDED USE
The thermometer is an electronic device intended to measure the body (human) temperature. It is designed to absorb the infrared from human body to detect your temperature. It can be used with adult or pediatric patients and it can be used in household.
CAUTION


| Temperature (°F) | Backlight |
| T ≤ 98.9 | Green Backlight |
| 99 ≤ T ≤ 99.4 | Yellow Backlight |
| 99.5 ≤ T | Red Backlight |
Install OUcare APP from Google play or App Store successfully. OUcare not only collect temperatures but also can do graph for checking temperature’s trend easily
For Android mobile device: When the device is off, put mobile device above the thermometer, the last temperature will be read-back and show on mobile device.
For iPhone7 or above: The iOS should be 11.0 or above. When the device is off, press the icon of APP then close to the thermometer, the last temperature will be read-back and show on mobile device. Right side is a figure of NFC area of iPhone7.
The thermometer is contraindicated:
Setting °C or °F Scale

Recall Memory

Delete All Memory
When the device is in memory mode, press and hold the power button to delete all memories, then the device will be off automatically.
Otherwise,



WARNING
Use only same type of alkaline batteries to avoid incompatibility.
Memory may be erased after replacing a new battery.

Replace Battery
Cleaning process:
Disinfection process:
Drying process:
Please contact your local agents or distributors if the device does not function properly as specified under Product Specification.
WARNING:
CAUTIONS:
ABOUT BODY TEMPERATURE
A person’s normal body temperature can range from 95.9°F to 100.0°F throughout the day. We recommend taking your temperature at different times during the day when you are NOT sick. This will give you a preliminary reference to know what your normal temperature is and how it fluctuates during a normal day. Then, when you feel sick or uncomfortable, knowing this information will provide an index for a reliable diagnosis of a fever.
Warranty:
The device itself (battery excluded) comes with one year warranty from purchase date if it is operated under normal circumstances without evidence of tampering. See exclusions below.
Warranty exclusions:
LOT NUMBER INFORMATION
LOT NO.(YYMM) on the device label means the year and month of manufacturing. For example, “1904” means the device was manufactured in April, 2019.

Thank you for purchasing a 180 Innovations product. Distributed by 180 Innovations, Lakewood, CO 80215 Made in China
www.180innovations.com