This device is intended for noninvasive spot-checking your blood oxygen saturation levels and pulse rate.
* Please refer to the manual for detailed operation of the Display Switching & Parameter Setting.
Section 1
Instructions for the Safe Operation and Use of the Fingertip Pulse Oximeter
Warnings
CAUTION
Batch code, Date of manufacturer and Serial No are printed on the label on the battery cover.
Section 2
General
This chapter provides a general description of the Fingertip Pulse Oximeter including:
Indication for use/ intended use
The Pulse Oximeter is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for sport and aviation use, not in clinical institution.
Brief Device Description
The Fingertip Pulse Oximeter, based on all digital technology, is intended for noninvasive spot-check measurement of functional oxygen saturation of arterial hemoglobin (SpO2). Advanced DSP algorithm* can minimize the influence of motion artifact and improve measurement accuracy of low perfusion*. The Oximeter can be used to measure human Hemoglobin Saturation and heart rate through finger. The product is suitable for family, sport, aviation and etc.
Product Features
CAUTION
PI (Perfusion Index )is the ratio of the pulsatile blood flow to the non-pulsatile static blood flow in a body’s peripheral tissue, such as finger tip, toe, or ear lobe. Perfusion index is an indication of the pulse strength at the sensor site.
Section 3
Description of the Front Panel
Part Definition and Description
Item | Name | Description |
1 | Power button | Turn on the machine |
2 | OLED Panel | Display the SPO2/PR data & Plethysmogram |
3 | Battery Compartment |
Display
After switch on, the OLED display of the Fingertip Pulse Oximeter is as follows:
Parameter setting
When the device is under measuring interface, press the select button for 1 second in order to enter into menu page (figure 3.3.1 and figure 3.3.2 ). There are two submenus for choice:
Remind Setup
Press the select button for 1 second and enter into the Remind Setup menu. User can adjust the setting through moving the “*” symbol to the back of the Sound Reminder, Beep, Restore or Brightness.
Limit Value Setting
When the * symbol show on the Reminder Setup, long press the select button until enter into the Remind Limit setup menu (figure 3.3.2). User can press the select button to select the items. And press the select button for 1 second to change the data you need.
On the Reminder Limit setup menu page (figure 3.3.2), when the * symbol show behind the “+/-”. Press select button for 1 second to change the “+” to “-” or change the “-” to “+”.
When “+” shows on the right side, move the “*” after the Spo2 Hi or PR Hi setting and long press the select button for 1 second can increase the value to a higher value (until it reaches to the highest.)
When “-” shows on the right side, move the “*” after the Spo2 Lo or PR Lo value setting and long press the select button for 1second can reduce the value to a lower value (untill it reaches to the lowest).
Note:
Install battery
Installing two AAA batteries into battery cassette in correct polarities and cover it.
WARNING: Do not attempt to recharge normal alkaline batteries, they may leak and may cause a fire or even explode.
Turn the Fingertip Pulse Oximeter on Put one of fingers into rubber hole of the Oximeter (it is best to put the finger thoroughly) with nail surface upward, then releasing the clamp.
Press power button to turn the Fingertip Pulse Oximeter on.Read correspondent data from display screen.
Display Description
of OLED It is shown as below:
Note:
Section 4
Cleaning
Switch off the power and take out the batteries before cleaning, Keep the exterior surface of the device clean and free of dust and dirt. Cleaning exterior surface (OLED display screen included) of the unit with a dry and soft cloth. Use 75% density of medical alcohol to clean the surface and use dry fabric with little alcohol to avoid alcohol permeates into the device.
Disinfection
Disinfecting the machine after using by mutiple users.
Use 75% density of medical alcohol to clean the surface that contacting with the users.
CAUTION
Section 5
Maintenance
Troubleshooting
Problems | Possible Reason | Resolutions |
Oxyhemo- globin or heart rate can not be shown normally | 1. Finger is not plugged correctly.
2. User’s perfusion is too low to be measured. |
1. Retry by plugging the finger
2. Try some more times, if you can make sure about no problem existing in the product, Please go to a hospital timely for exact diagnosis |
Oxyhemo- globin or heart rate is shown unstably | 1. Finger might not be plugged deep enough
2. Finger is trembling or user’s body is in movement status |
1. Retry by plugging the finger
2. Try not to move, let the user keep calm |
Oxyhemo- globin or heart rate is abnormal and cause sound reminder | 1. Finger is not plugged correctly.
2. User’s SPO2&PR is abnormal. |
1. Retry by plugging the finger
2. go to the hospital for further examination |
The oximeter can not be powered on | 1. Power of batteries might be inadequate or not be there at all 2.Batteries might be installed incorrectly 3.The Oximeter might be damaged | 1.Please replace batteries 2.Please reinstall the batteries
3.Please contact with local customer service center |
The screen are suddenly off | 1.The product is automa- tically powered off when no signal is detected longer than 16 seconds 2.Power quantity of the batteries is exhausted | 1.Normal
2.Replace the batteries |
Section 6
Fingertip Pulse Oximeter Specifications: Physical Characteristics
Machine:
Dimensions :
62 mm (L) x 34mm (W) x 31mm (D)
Weight approx: 50g
(including 2 x AAA battery)
Classification :
Anti-electric Shock Type:
Internally powered equipment
Anti-electric Shock Degree:Type BF equipment
EMC:Type B
Mode of operation: Continuous Operation Enclosure Degree of ingress protection: IP22
IP22 means shell of this product can withstand the water dropping to the surface when the shell deviate 15 degree from horizontal surface.
Power
Internal | 2 x AAA Ikaline battery |
Power consumption | Samller then 30mA (normal) |
Manufacturer’s Declaration of the EMC
Guidance and manufacturer´s declaration – electromagnetic emission –for all EQUIPMENT AND SYSTEMS
1 | Guidance and manufacturer´s declaration – electromagnetic emission | ||
2 | The Pulse Oximeter is intended for use in the electromagnetic environment specified below.
The customer or the user of the Pulse Oximeter should assure that it is used in such an environment. |
||
3 | Emissions test | Compliance | Electromagnetic environment – guidance |
4 | RF emissions
CISPR 11 |
Group 1 | The Pulse Oximeteruses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. |
5 | RF emissions
CISPR 11 |
Class B | The Pulse Oximeter is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. |
6 | Harmonic emissions IEC 61000-3-2 | N/A | |
7 | Voltage fluctuations /
flicker emissions IEC 61000-3-3 |
N/A |
Guidance and manufacturer’s declaration – electromagnetic immunity –for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of the Pulse Oximetershould assure that it is used in such an environment.
Immunity test | IEC 60601
test level |
Compliance level | Electromagnetic environment guidance |
Electrostatic discharge (ESD)
IEC 61000-4-2 |
± 8 kV contact
± 15 kV air |
± 8 kV contact
± 15 kV air |
Floors should be wood, concrete orceramic tile. If floors are covered withsynthetic material, the relative humidityshould be at least 30 %. |
Electrostatic | ± 2 kV for power | ||
transient / burst
IEC 61000-4-4 |
supply lines
± 1 kV for input/ outputlines |
N/A | Mains power quality should be that of a typical commercial or hospital environment. |
Surge
IEC 61000-4-5 |
± 1 kV differential mode
± 2 kV common mode |
N/A | Mains power quality should be that of a typical commercial or hospital environment. |
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11 |
0 % UT; 0,5 cycle g)
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315° 0 % UT; 1 cycle and 70 % UT; 25/30 cycles at 0° 0 % UT; 250/300 cyclee |
N/A | Mains power quality should be that of a typical commercial or hospital environment. If the user of the Pulse Oximeter requires continued operation during power mains interruptions, it is recommended that the Pulse Oximeter
be powered from an uninterruptible power supply or a battery. |
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8 |
30A/m | 30A/m | Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. |
NOTE UT is the a. c. mains voltage prior to application of the test level. |
Guidance and manufacturer´s declaration – electromagnetic immunity –for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Guidance and manufacturer´s declaration – electromagnetic immunity | |||
The Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of the Pulse Oximeter should assure that it is used in such an environment. | |||
Immunity test | IEC 60601
test level |
Compliance level | Electromagnetic environment – guidance |
Conducted RF | 3 Vrms | N/A | Portable and mobile RF communications equipment should be used no closer to any part of the Pulse Oximeter, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance d = [3.5] P d = [3.5] P 80 MHz to 800 MHz V 1 E1 d = [ 7 ] P 800 MHz to 2.5 GHz E1 where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).b Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. b Interference may occur in the vicinity of equipment marked with the following symbol: |
IEC 61000-4-6 | 150 kHz to
80 MHz |
||
Radiated RF | 6Vrms in ISM banda between 150 kHz to
80 MHz |
10 V/m | |
IEC 61000-4-3 | 80 MHz to
2.7 GHz |
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Pulse Oximeter is used exceeds the applicable RF compliance level above, the Pulse Oximeter should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Pulse Oximeter.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM -for EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the Pulse Oximeter
The Pulse Oximeter is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Pulse Oximeter can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Pulse Oximeter as recommended below, according to the maximum output power of the communications equipment
Rated maximum output of transmitter W | Separation distance according to frequency of transmitter / m | ||
150 kHz to
80 MHz d = [3.5] P V 1 |
80 MHz to
800 MHz d = [3.5] P E1 |
800 MHz
to 2.7 GHz d = [ 7 ] P E1 |
|
0.01 | / | 0.12 | 0.23 |
0.1 | / | 0.38 | 0.73 |
1 | / | 1.2 | 2.3 |
10 | / | 3.8 | 7.3 |
100 | / | 12 | 23 |
For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. |
Shenzhen Aeon Technology Co., Ltd. RM6H02, Block 27-29 Tianxia IC Industrial Park, Majialong No.133 of Yiyuan road, Nantou Street Nanshan District, Shenzhen, China.
Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537 Hamburg, Germany